Enforcement Report - Week of November 7, 2012
-
Back to Previous Page
Product Detail
| Product Description | Myocardial Heart Wires Intended for temporary atrial and ventricular pacing and sensing during or after cardiac surgery in conjunction with an external pulse generator. |
|---|---|
| Recall Number | Z-0165-2013 |
| Classification | Class II |
| Code Info | Model Number TME60S, Lot #C1-07101 Model Number TME64S, Lot #C1-07056 Model Number TME64S, Lot #C1-07057 Model Number TME64S, Lot #C1-07058 Model Number TME64S, Lot #C1-07111 Model Number TME64S, Lot #C1-07112 Model Number TME64S, Lot #C1-07113 Model Number TME64S, Lot #C1-07114 Model Number TME64S, Lot #C1-07122 Model Number TME64S, Lot #C1-07123 Model Number TME64S, Lot #C1-07124 Model Number TME64S, Lot #C1-07125 Model Number TME64C, Lot #C1-07143 Model Number TME64S, Lot #C1-07151 Model Number TME64S-3, Lot #C1-07156 Model Number TME65S, Lot #C1-07174 |
| Product Distributed Qty | 1425 |
| Reason For Recall | Oscor, Inc. in Palm Harbor, FL is recalling myocardial heart wires, models TME60S, TME64S, TME64C, TME64S-3, TME65S, and TME66S. The recall was initiated due to the spacing between anchoring zig-zag to electrode is out of specification. |
Event Detail
| Event Id | 63112 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Oscor, Inc. |
| City | Palm Harbor |
| State | FL | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-07-25 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA including OR, FL, Al, TX, NM, CA, AL, PA, KY, IN, OH, and IL and internationally to Canada, Russia, and New Zealand |
-