Product Detail

Product Description Plum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
Recall Number Z-0203-2014
Classification Class II
Code Info List No. 11005; Serial Numbers: 16792201, 16792202, 16792205, 16792206, 16792208, 16792209, 16792210, 16792211, 16792213, 16792216, 16792217, 16792219, 16792220, 16792222, 16792223, 16792224, 16792225, 16792226, 16792227, 16792228, 16792230, 16792232, 16792233, 16792234, 16792235, 16792236, 16792238, 16792239, 16792240, 16792241, 16792242, 16792244, 16792245, 16792246, 16792248, 16792249, 16792250, 16792251, 16792252, 16792253, 16792254, 16792255, 16792256, 16792257, 16792258, 16792259, 16792260, 16792261, 16792263, 16792264, 16792265, 16792266, 16792268, 16792269, 16792271, 16792272, 16792273, 18807402, 18807405, 18807406, 18807407, 18807408, 18807409, 18807410, 18807413, 18807414, 18807417, 18807418, 18807420, 18807421, 18807423, 18807427, 18807428, 18807429, 18807430, 18807431, 18807432, 18807433, 18807434, 18807439, 18807443, 18807444, 18807446, 18807449, 18807452, 18807453, 18807455, 18807456, 18807459, 18807463, 18807466, 18807472, 18807474, 18807478, 18807484, 18807485, 18807486, 18807490, 18807491, 18807494, 18807496, 18807499, 18807500, 18807501, 18807506, 18807509, 18807510, 18807511, 18807513, 18807514, 18807518, 18807523, 18807524, 18807529, 18807535, 18807536, 18807539, 18807540, 18807541, 18807542, 18807543, 18807545, 18807546, 18807556, 18807557, 18807568, 18807574, 18807575, 18807576, 18807578, 18807581, 18807584, 18807585, 18807586, 18807588, 18807589, 18807590, 18807591, 18807596, 18807597, 18807598, 18807599, 18807600, 18807601, 18807602, 18807604, 18807605, 18807607, 18807608, 18807609, 18807610, 18807611, 18807612, 18807613, 18807614, 18807617, 18807618, 18807619, 18807620, 18807623, 18807625, 18807626, 18807627, 18807628, 18807630, 18807631, 18807632, 18807633, 18807634, 18807636, 18807637, 18807639, 18807640, 18807695, 18807696, 18807697, 18807698, 18807699, 18807700
Product Distributed Qty List No. 11005: 179 units
Reason For Recall Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not be detected. These errors invoke visual and audible warnings to the user. Hospira has identified the following design attributes of the pressure sensor that may cause the calibration to drift sooner than expected: strain gage protective layers' moisture resistance issues, pressure assembly chassis residual structural stress issues, and relaxation of clamping force from mounting screws for strain gage and pin stack.

Event Detail

Event Id 64091
Product Type Devices
Status Ongoing
Recalling Firm Hospira Inc.
City Lake Forest
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-02-05
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Croatia, Dominican Republic, Turks and Caicos.
-
-