• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of November 14, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
Recall Number Z-0256-2013
Classification Class II
Code Info catalog ##053-2311-4, all serial numbers
Product Distributed Qty 69 units
Reason For Recall There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Event Detail

Event Id 62580
Product Type Devices
Status Ongoing
Recalling Firm Sysmex America, Inc.
City Mundelein
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-01
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide distribution: USA including Puerto Rico.
-
-