Enforcement Report - Week of November 14, 2012
-
Back to Previous Page
Product Detail
| Product Description | Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem. |
|---|---|
| Recall Number | Z-0267-2013 |
| Classification | Class II |
| Code Info | Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243) |
| Product Distributed Qty | 12 units |
| Reason For Recall | Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches. |
Event Detail
| Event Id | 53337 |
|---|---|
| Product Type | Devices |
| Status | Completed |
| Recalling Firm | Medacta Usa Inc |
| City | Camarillo |
| State | CA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2008-10-13 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium. |
-