Product Detail

Product Description S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
Recall Number Z-0271-2013
Classification Class II
Code Info Serial Numbers: 10E06219,10E06220, 10E06231-10E06234, 10E06239-10E06241, 10E06255-10E06273, 10E06286-10E06289, 10E06291-10E06355, 10E06359-10E06364, 10E06366, 10E06368-10E06383, 10E06385, 10E06387-10E06391,10E06395-10E06401, 10E06403-10E06447, 10E6502.
Product Distributed Qty 35 units
Reason For Recall Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determined to be the cause of some of these events.

Event Detail

Event Id 63383
Product Type Devices
Status Ongoing
Recalling Firm Sorin Group USA, Inc.
City Arvada
State CO
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-09-19
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.
-
-