Product Detail

Product Description PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Recall Number Z-0287-2013
Classification Class II
Code Info Catalog #: PT-Module
Product Distributed Qty 136 units
Reason For Recall Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.

Event Detail

Event Id 53336
Product Type Devices
Status Terminated
Recalling Firm Dako North America Inc.
City Carpinteria
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2008-05-01
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.
-
-