Product Detail

Product Description The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+ Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual. The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
Recall Number Z-0290-2014
Classification Class II
Code Info Excessive data in the Code Information for this product has caused issues with the CSV download function. The complete list of Code Information can be found at the <a href="http://www.fda.gov/Safety/Recalls/EnforcementReports/UCM376115" target="_blank">Addendum to Enforcement Report</a>.
Product Distributed Qty 1) List Number 20679: 7,737 units; 2) List Number 20792:
Reason For Recall Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken distal pressure pin could result in incorrect distal pressure readings, undetected distal occlusions and/or undetected cassette failures.

Event Detail

Event Id 64263
Product Type Devices
Status Ongoing
Recalling Firm Hospira Inc.
City Lake Forest
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-02-01
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Costa Rica and Dominican Republic.
-
-