Enforcement Report - Week of November 28, 2012
-
Back to Previous Page
Product Detail
| Product Description | Product is Full OSSEOTITE 2 Tapered Certain Implant, Model Number XIFNT413, and size 4 x 13 mm. The product is sterile. Product provides a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures. |
|---|---|
| Recall Number | Z-0297-2013 |
| Classification | Class II |
| Code Info | Model Number XIFNT413, Lot # 2012040462 |
| Product Distributed Qty | 191 devices |
| Reason For Recall | Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This condition may prevent the driver or abutment from fully engaging in the implant. |
Event Detail
| Event Id | 63618 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Biomet 3i, LLC |
| City | Palm Beach Gardens |
| State | FL | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-09-28 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution-USA (nationwide) including the states of VA, SC, HI, LA, OH, FL, IL, NJ, NC, CA, CT, MI, TN, AK, NH, NY, MA, GA, PA, and KY and the countries of United Kingdom, Spain, Germany, Canada, Japan, Singapore, Taiwan, Israel, Poland, Hong Kong, Ireland, The Netherlands, Austria, and France. |
-







