• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of November 28, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile. Product provides a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.
Recall Number Z-0298-2013
Classification Class II
Code Info Model Number IFOA411, Lot # 2011111333, 2011111333-S1
Product Distributed Qty 76 devices
Reason For Recall Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This condition may prevent the driver or abutment from fully engaging in the implant.

Event Detail

Event Id 63618
Product Type Devices
Status Ongoing
Recalling Firm Biomet 3i, LLC
City Palm Beach Gardens
State FL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-09-28
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) including the states of VA, SC, HI, LA, OH, FL, IL, NJ, NC, CA, CT, MI, TN, AK, NH, NY, MA, GA, PA, and KY and the countries of United Kingdom, Spain, Germany, Canada, Japan, Singapore, Taiwan, Israel, Poland, Hong Kong, Ireland, The Netherlands, Austria, and France.
-
-