• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of November 28, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description SenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use with the V-Sign Sensor only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.
Recall Number Z-0433-2013
Classification Class II
Code Info Lot numbers: 1CF5, 1D21, 1D60, 1DC6, 1DED, 1E0D
Product Distributed Qty 195 boxes (9 per box)
Reason For Recall A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

Event Detail

Event Id 63631
Product Type Devices
Status Ongoing
Recalling Firm Sentec AG
City Therwil, Baselland
State
Country CH
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-05
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Distributed only in MO.
-
-