• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of December 11, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.
Recall Number Z-0433-2014
Classification Class II
Code Info Siemens Material Number 10800057 (Version VA11B) and 10803188 (Version VA12A)
Product Distributed Qty 262
Reason For Recall Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.

Event Detail

Event Id 65147
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc.
City Newark
State DE
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-10
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution: USA (nationwide) and internationally to Canada
-
-