Enforcement Report - Week of December 11, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description SIEMENS brand ONCOR series Linear Accelerator Linac systems with component: any photon unflat beam option(Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, Roentgenstrasse 19-21, Kemnath, GERMANY 95478 and distributed SIEMENS Healthcare IM CR RO, CR  Radiation Oncology, Martinez, CA.
Recall Number Z-0437-2014
Classification Class II
Code Info Serial Numbers: 5281, 5385, 5445
Product Distributed Qty 3 units
Reason For Recall SIEMENS became aware of two (2) customer complaints concerning a safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.

Event Detail

Event Id 66724
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Martinez
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-10-23
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.
-
-