Enforcement Report - Week of December 19, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the quantitative determination of theophylline in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.
Recall Number Z-0512-2013
Classification Class II
Code Info All lot codes
Product Distributed Qty 50 test=2397 and 250 test= 594
Reason For Recall Operating range of the ADVIA Centaur systems is 18°C to 30°C (64.4°F to 86.0°F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Event Detail

Event Id 63682
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc
City East Walpole
State MA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-14
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
-
-