Product Detail

Product Description FastPack® Testosterone Immunoassay Chemilunescence assay for the determination of Testosterone Testosterone The FastPack® Testo Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of total testosterone in human serum and plasma. It is intended strictly for in-vitro diagnostic use as an aid in the diagnosis and management of conditions involving excess or deficiency of this androgen. The FastPack® Testo Immunoassay is designed for use with the FastPack® System.
Recall Number Z-0548-2014
Classification Class II
Code Info Testosterone. Catalog Number 25000041. Lot Number 1310019-3P Exp Date: 9/22/2014
Product Distributed Qty 18
Reason For Recall Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.

Event Detail

Event Id 66842
Product Type Devices
Status Ongoing
Recalling Firm Qualigen Inc
City Carlsbad
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-11-08
Initial Firm Notification of Consignee or Public Telephone
Distribution Pattern US Distribution to states of: AL, CA, KS, KY, IN, MA, NC, NY, NV, OK, SC, TN, TX, UT, VA, and WA.
-
-