• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of December 26, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description FastPack® Free T4 Immunoassay Chemilunescence assay for the determination of Free T4 Free T4 The FastPack® free T4 Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of FT4 in human serum. The FastPack® free T4 Immunoassay is designed for use with the FastPack® System.
Recall Number Z-0549-2014
Classification Class II
Code Info Testosterone. Catalog Number 25000043. Lot Number 1310028-2P.Exp Date: 4/10/2014
Product Distributed Qty 17
Reason For Recall Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.

Event Detail

Event Id 66842
Product Type Devices
Status Ongoing
Recalling Firm Qualigen Inc
City Carlsbad
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-11-08
Initial Firm Notification of Consignee or Public Telephone
Distribution Pattern US Distribution to states of: AL, CA, KS, KY, IN, MA, NC, NY, NV, OK, SC, TN, TX, UT, VA, and WA.
-
-