Product Description Ziehm Vision C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic Imaging
Recall Number Z-0564-2014
Classification Class II
Code Info 9209 9277 9448 9584 9655 9211 9278 9459 9585 9656 9212 9279 9475 9586 9657 9214 9315 9478 9587 9658 9216 9316 9506 9588 9659 9217 9343 9514 9633 9660 9245 9352 9530 9634 9664 9246 9358 9531 9645 9665 9255 9376 9551 9646 9666 9256 9377 9562 9647 9672 9272 9404 9571 9653 97119273 9427 9582 9654 9728
Product Distributed Qty 60 devices
Reason For Recall Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.

Event Detail

Event Id 66976
Product Type Devices
Status Ongoing
Recalling Firm Ziehm Imaging Inc
City Orlando
State FL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-11-21
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern USA (Nationwide) Distribution including the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MT, MA, MD, MI, MN, MN, MO,MS, NC ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Hawaii.
-
-