Product Detail
| Product Description | Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers. |
|---|---|
| Recall Number | Z-0608-2013 |
| Classification | Class III |
| Code Info | Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008. |
| Product Distributed Qty | 840 Ampules |
| Reason For Recall | The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45. |
Event Detail
| Event Id | 37530 |
|---|---|
| Product Type | Devices |
| Status | Completed |
| Recalling Firm | Radiometer America Inc |
| City | Westlake |
| State | OH | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2006-06-06 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Distributed to KY, NC, and NY. |
-