Product Detail

Product Description Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.
Recall Number Z-0608-2013
Classification Class III
Code Info Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.
Product Distributed Qty 840 Ampules
Reason For Recall The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Event Detail

Event Id 37530
Product Type Devices
Status Completed
Recalling Firm Radiometer America Inc
City Westlake
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2006-06-06
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Distributed to KY, NC, and NY.
-
-