Enforcement Report - Week of January 2, 2013
-
Back to Previous Page
Product Detail
| Product Description | Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron Access Clinical System Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer. |
|---|---|
| Recall Number | Z-0612-2013 |
| Classification | Class II |
| Code Info | Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640 |
| Product Distributed Qty | 6,858 units |
| Reason For Recall | It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel. |
Event Detail
| Event Id | 63632 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Beckman Coulter Inc. |
| City | Brea |
| State | CA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-11-01 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution-USA (nationwide) and the country of Canada. |
-