Enforcement Report - Week of January 2, 2013
-
Back to Previous Page
Product Detail
| Product Description | COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland |
|---|---|
| Recall Number | Z-0614-2013 |
| Classification | Class II |
| Code Info | CAP 394436 to 394663 |
| Product Distributed Qty | 7 units |
| Reason For Recall | Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots. |
Event Detail
| Event Id | 63814 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Roche Molecular Systems, Inc. |
| City | Branchburg |
| State | NJ | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-10-11 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Distributed in the states of GA, CA, NC, and MN. |
-