Enforcement Report - Week of January 9, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971; b) list 11973 - software version 10.3; c) list 12391 - software version 11.3; d) list 20679 - Hospira MedNet Software e) list 20792 - Hospira MedNet Software; Product Usage: The Plum A+ Infusion System is a dual-line volumetric infusion pump that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care
Recall Number Z-0624-2013
Classification Class II
Code Info list numbers 11971, 11973, 12391, 20679, and 20792; all serial numbers
Product Distributed Qty 297,833 units
Reason For Recall The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.

Event Detail

Event Id 63772
Product Type Devices
Status Ongoing
Recalling Firm Hospira Inc.
City Lake Forest
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-12-07
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.
-
-