Enforcement Report - Week of January 9, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software; Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care
Recall Number Z-0625-2013
Classification Class II
Code Info list numbers 12348, 12618, and 20678; all serial numbers
Product Distributed Qty 28,599 units
Reason For Recall The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.

Event Detail

Event Id 63772
Product Type Devices
Status Ongoing
Recalling Firm Hospira Inc.
City Lake Forest
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-12-07
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.
-
-