Enforcement Report - Week of January 9, 2013
-
Back to Previous Page
Product Detail
| Product Description | Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls |
|---|---|
| Recall Number | Z-0628-2013 |
| Classification | Class II |
| Code Info | Model 3002SEX, Serial Number 120315371-120315379 |
| Product Distributed Qty | 9 |
| Reason For Recall | Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes. |
Event Detail
| Event Id | 63602 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Stryker Medical Division of Stryker Corporation |
| City | Portage |
| State | MI | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-12-18 |
| Initial Firm Notification of Consignee or Public | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
| Distribution Pattern | Worldwide Distribution - Nationwide and Canada |
-