• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of January 9, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description ABX PENTRA AST CP, Part Number: A11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in vitro determination of Aspartate Amino Transferase (AST) in serum or plasma by colorimetry.
Recall Number Z-0632-2013
Classification Class II
Code Info Lot number: 00456-4108
Product Distributed Qty 66886 units
Reason For Recall Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.

Event Detail

Event Id 63838
Product Type Devices
Status Ongoing
Recalling Firm Horiba Instruments, Inc dba Horiba Medical
City Irvine
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-10
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX.
-
-