Enforcement Report - Week of January 9, 2013
-
Back to Previous Page
Product Detail
| Product Description | ABX PENTRA AST CP, Part Number: A11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in vitro determination of Aspartate Amino Transferase (AST) in serum or plasma by colorimetry. |
|---|---|
| Recall Number | Z-0632-2013 |
| Classification | Class II |
| Code Info | Lot number: 00456-4108 |
| Product Distributed Qty | 66886 units |
| Reason For Recall | Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm. |
Event Detail
| Event Id | 63838 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Horiba Instruments, Inc dba Horiba Medical |
| City | Irvine |
| State | CA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-10-10 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX. |
-