Enforcement Report - Week of January 16, 2013
-
Back to Previous Page
Product Detail
| Product Description | Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 When used with the CLINITEK ATLAS® Automated urine Chemistry Analyzer and CLINITEK ATLAS® 10 reagent Pak, CLINITEK ATLAS® Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas. |
|---|---|
| Recall Number | Z-0636-2013 |
| Classification | Class III |
| Code Info | Carton/Bottle Lot Number: S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S0100101B P0100101B S0103042A P0103042A S0103042H P0103042H S0106062A P0106062A S9101022A P9101022A S9101022B P9101022B S9105052A P9105052A |
| Product Distributed Qty | 6466 |
| Reason For Recall | When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F). |
Event Detail
| Event Id | 63809 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Healthcare Diagnostics Inc |
| City | Norwood |
| State | MA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-11-21 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong, |
-