Product Detail

Product Description Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 When used with the CLINITEK ATLAS® Automated urine Chemistry Analyzer and CLINITEK ATLAS® 10 reagent Pak, CLINITEK ATLAS® Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas.
Recall Number Z-0636-2013
Classification Class III
Code Info Carton/Bottle Lot Number: S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S0100101B P0100101B S0103042A P0103042A S0103042H P0103042H S0106062A P0106062A S9101022A P9101022A S9101022B P9101022B S9105052A P9105052A
Product Distributed Qty 6466
Reason For Recall When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25°C (77°F).

Event Detail

Event Id 63809
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics Inc
City Norwood
State MA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-21
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
-
-