Enforcement Report - Week of January 16, 2013
-
Back to Previous Page
Product Detail
| Product Description | Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for use in the quantitative analysis of Phenobarbital in human serum or plasma. |
|---|---|
| Recall Number | Z-0664-2013 |
| Classification | Class II |
| Code Info | Emit(R) 2000 Phenobarbital Assay (OSR4D229) lots 1237 and 1264 and/or Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots E1, E2, and E3. |
| Product Distributed Qty | 1979 cartons |
| Reason For Recall | Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an application if the AU(R) System is used, the issue may be experienced. Firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. This is an expansion of the recall initiated in July, 2012 to include additional lots. |
Event Detail
| Event Id | 63892 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
| City | Newark |
| State | DE | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-12-10 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Distributed in the states of CA, FL, IL, KS, MA, MI, MN, MS, NM, NY, NC, OH, OR, PA, TN, TX, UT, and WA. |
-