Product Detail

Product Description Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.
Recall Number Z-0675-2013
Classification Class II
Code Info Model #728269; Serial #s: 79906, 79921, 79927, 79937, 80007, 80026, 80031, 80040, 80058, 80062, 80066, 80069, 80087, 80110, 80121, 80155, 80167, 80168, 80169, 80174, 80192, 80193, 80195, 80201, 80208, 80209, 80219, 80227, 80228, 80230, 80252, 80258, 80266, 80280, 80289, 80293, 80294, 80315, 80322, 80328, 80333, 80340, 80341, 80348, 80360, 80375, 80385, 80387, 80394, 80398, 80416, 80421, 80422, 80429, 80447, 80451, 80460, 80474, 80482, 80484, 80485, 80492, 80497, 80505, 80516, 80578, 80585, 80591, 80629, 80646, 80658, 80681, 80704, 80709, 80711, 80720, 80723, 80743, 80749, 80751, 80831, 80835, 80866, 80878, 80908, 80909, 80911, 80914, 80916, 80930, 80931, 80944, 80974, 80992, 80999, 81023, 81025, 81034, 81045, 81071, 81099, 81101, 81107, 81117, 81126, 81136, 95382 & 95592.
Product Distributed Qty 108 units
Reason For Recall Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image are segmented automatically. However, the center line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu.

Event Detail

Event Id 63393
Product Type Devices
Status Ongoing
Recalling Firm Philips Medical Systems (Cleveland) Inc
City Cleveland
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-09-11
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution-USA including the states of AZ, CA, CO, IL, IL, IN, MA, ME, MI, MO, NC, NJ, NY, OR, TN, TX, VA and WI and the countries of Austria, Canada, China, France, India, Ireland, Italy, Japan, Netherlands, Norway, Spain, Switzerland, Thailand and United Kingdom.
-
-