Enforcement Report - Week of January 23, 2013
-
Back to Previous Page
Product Detail
| Product Description | DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct. |
|---|---|
| Recall Number | Z-0685-2013 |
| Classification | Class II |
| Code Info | All lots |
| Product Distributed Qty | 140 units |
| Reason For Recall | A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification. |
Event Detail
| Event Id | 62719 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | DePuy Spine, Inc. |
| City | Raynham |
| State | MA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-07-24 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK. |
-







