Enforcement Report - Week of January 30, 2013
-
Back to Previous Page
Product Detail
| Product Description | DePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component. |
|---|---|
| Recall Number | Z-0710-2013 |
| Classification | Class II |
| Code Info | All lots |
| Product Distributed Qty | 23 units |
| Reason For Recall | Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPERĀ® large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head |
Event Detail
| Event Id | 63665 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | DePuy Spine, Inc. |
| City | Raynham |
| State | MA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-11-07 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia. |
-