Enforcement Report - Week of February 6, 2013
-
Back to Previous Page
Product Detail
| Product Description | Cordis, a Johnson and Johnson Company, ExoSeal(TM) 6F Vascular Closure Device REF cat. No EX500, EX600, and EX700. Used for femoral artery puncture site closure, reducing times to hemostasis and ambulation in patients who have undergone diagnostic or interventional procedures. |
|---|---|
| Recall Number | Z-0740-2013 |
| Classification | Class II |
| Code Info | Model EX500: Lot #'s: 15410906, 15411659, 15424968, 15429098, 15429879, 15437509, 15437958, 15444208, 15449422, 15453540, 15456567, 15456568, 15462779, 15462780, 15469984, 15479199, 15481341, 15485491, 15489970, 15501955, 15505186, 15578709, 15589368, 15586530, 15602199, 15603369, 15608590, 15610801, 15617548, 15630141, 15630171, 15639388, 15642692, 15648662, 15650923, 15656441, 15664083, 15670351, 15670768. Model EX600: Lot #'s: 15408606, 15409332, 15410622, 15410623, 15426890, 15430798, 15433095, 15433098, 15441594, 15443327, 15446063, 15448005, 15451542, 15452987, 15455770, 15455771, 15459366, 15459367, 15461179, 15461781, 15463296, 15463297, 15464298, 15464299, 15466362, 15472044, 15475128, 15475795, 15479644, 15484135, 15487458, 15491737, 15496311, 15497038, 15500078, 15500709, 15503015, 15504223, 15504224, 15504748, 15506926, 15508012, 15508353, 15508856, 15510541, 15512713, 15518232, 15540529, 15545067, 15548555, 15549324, 15550196, 15550197, 15557284, 15561545, 15561546, 15566744, 15572185, 15573480, 15579468, 15585121, 15587184, 15596793, 15601234, 15601753, 15604477, 15605089, 15609335, 15613282, 15617549, 15625033, 15627316, 15635233, 15637262, 15643624, 15644116, 15644789, 15644790, 15649759, 15663330. Model EX700 Lot #'s: 15466361, 15471271, 15471272, 15476317, 15476743, 15482155, 15483371, 15486224, 15490397, 15535011, 15540530. |
| Product Distributed Qty | 98,675 units distributed in US. |
| Reason For Recall | On 10/29/2012 Cordis Corporation, Miami Lakes, FL initiated a recall of their Cordis ExoSeal (TM) Vascular Closure Device. Affected lots of Cordis ExoSeal (TM) VCD have been sterilized using a process that did not achieve the sterility assurance level that is required by the standards of this class of product. Model Numbers EX500, EX600, and EX700. |
Event Detail
| Event Id | 63827 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Cordis Corporation |
| City | Miami Lakes |
| State | FL | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-10-29 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | US (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Hawaii. |
-