Product Detail

Product Description MCA5-05SD - 0.5 MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.
Recall Number Z-0750-2014
Classification Class II
Code Info All lots
Product Distributed Qty 212
Reason For Recall The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decided to recall the cutting burrs along with the dissection tools.

Event Detail

Event Id 63624
Product Type Devices
Status Ongoing
Recalling Firm The Anspach Effort, Inc.
City Palm Beach Gardens
State FL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-03-29
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.
-
-