Product Detail

Product Description CareFusion SmartSite Gravity Blood Set Model 42082E. Used to administer blood and blood products through a needle or catheter inserted into the patient's artery or vein.
Recall Number Z-0832-2013
Classification Class II
Code Info Lot number 12055225
Product Distributed Qty 5,050 units
Reason For Recall CareFusion is recalling the SmartSite® Gravity Blood Set as a result of a potential for separation at the connection of the drip chamber and tubing . The separation may be observed at priming or observed during use. If a separation between the drip chamber and tubing occurs, this could cause a delay in treatment.

Event Detail

Event Id 64204
Product Type Devices
Status Ongoing
Recalling Firm Carefusion 303 Inc
City San Diego
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-12-20
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution including the states of AK, CA, CO, IL, MO, NV, NY, TX and UT.
-
-