Product Detail

Product Description Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium.
Recall Number Z-0913-2013
Classification Class II
Code Info All serial numbers / Multiple manufacture dates. Part# Item# ================ A94907 AU5811-02 A94912 AU5821-02 A94917 AU5831-02 A94922 AU5841-02 A94908 AU5811-03 A94913 AU5821-03 A94918 AU5831-03 A94923 AU5841-03 A94909 AU5811-04 A94914 AU5821-04 A94919 AU5831-04 A94924 AU5841-04 A94910 AU5811-06 A94915 AU5821-06 A94920 AU5831-06 A94925 AU5841-06
Product Distributed Qty 28 in U.S.; 4 Canada; 318 to other foreign countries.
Reason For Recall Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.

Event Detail

Event Id 64000
Product Type Devices
Status Ongoing
Recalling Firm Beckman Coulter Inc.
City Brea
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-28
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Austria, Brazil, China, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Korea Republic, Malaysia, Netherlands, Russian Federation, Spain, Switzerland, Taiwan, Turkey, United Kingdom, and Viet Nam.
-
-