Product Detail

Product Description REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument
Recall Number Z-0925-2013
Classification Class II
Code Info Catalog Number 5313152 Lot 153880,153930, 415500, 648010
Product Distributed Qty 58
Reason For Recall The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

Event Detail

Event Id 64354
Product Type Devices
Status Ongoing
Recalling Firm Biomet, Inc.
City Warsaw
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-02-05
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
-
-