Product Description Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.
Recall Number Z-0933-2013
Classification Class III
Code Info Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17
Product Distributed Qty 19, 503 cartons
Reason For Recall The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).

Event Detail

Event Id 64063
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc.
City Newark
State DE
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-12-20
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution
-
-