• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of April 3, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
Recall Number Z-0975-2013
Classification Class II
Code Info Part #: IFS-072-17-C, Lot #: 2399041; Part #: IFS-72-20-C, Lot #: 2399051 & Lot #: 2399051R and Part #: IFS-072-25-C, Lot #: 2399061.
Product Distributed Qty Lot # 2399041, 97 pcs; Lot # 2399051, 143 pcs; Lot # 2399051R, 4 pcs & Lot # 2399061, 119 pcs
Reason For Recall The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.

Event Detail

Event Id 64485
Product Type Devices
Status Ongoing
Recalling Firm Orthohelix Surgical Designs Inc
City Medina
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-10-30
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution including the states of CA, CO, IL, MI, MD, NC, NM, OH, OK, PA, SD, TX, UT, VA, WA and WI.
-
-