• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of April 3, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). Slide preparations serving as quality control challenges to demonstrate satisfactory performance of staining reagents.
Recall Number Z-1021-2013
Classification Class II
Code Info SL42-10 Lots 4239 to 4243 SL43-10 Lots 4333 to 4337
Product Distributed Qty 396 boxes of 10 slides each.
Reason For Recall Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control organisms are not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could potentially lead to misdiagnosis and improper treatment of a patient. There have been no complaints or reports of death or injury due to this issue.

Event Detail

Event Id 64469
Product Type Devices
Status Ongoing
Recalling Firm Microbiologics Inc
City Saint Cloud
State MN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-02-11
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution-USA including the states of CA, IN, MS, NY, OH, PA, and SD, and the countries of AUSTRALIA, AUSTRIA, UNITED ARAB EMIRATES, MALAYSIA, TAIWAN, JAPAN, TURKEY, BELGIUM, COLOMBIA, CANADA, SOUTH AFRICA, and VENEZUELA.
-
-