Product Detail

Product Description 20-Pole Eco Cable, Catalog #EM-5050-060 Product Usage: The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System.
Recall Number Z-1023-2013
Classification Class II
Code Info Impacted serial Numbers: 120, 132, 134,136, 151, 342, 350, 354, 356, 363, 368, 373, 379, 380, 384, 385, 402, 405, 424, 919, 953, 954, 960, 963, 968, 970, 977, 980, 993, 998, 1001, 1123, 1148, 1378, 1596, 1729, 1745, 1747, 1749, 1750, 1751, 1753, 1754, 1758, 1759, 1760, 1761, 1797, 1799, 1802, 1803, 1804, 1805. Lot Numbers: P1210, P1220, P1221, P1231.
Product Distributed Qty 54 units
Reason For Recall Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field.

Event Detail

Event Id 64579
Product Type Devices
Status Ongoing
Recalling Firm Biosense Webster, Inc.
City Irwindale
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-05
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (Nationwide) and countries of: DE, NL, CH, FR, GB, DK.
-
-