Product Detail

Product Description CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Recall Number Z-1114-2013
Classification Class II
Code Info catalog number 05530199160
Product Distributed Qty 4467 both devices
Reason For Recall Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.

Event Detail

Event Id 64460
Product Type Devices
Status Ongoing
Recalling Firm Roche Diagnostics Operations, Inc.
City Indianapolis
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-06
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution.
-
-