Product Detail

Product Description REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
Recall Number Z-1124-2013
Classification Class II
Code Info 32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103
Product Distributed Qty 430
Reason For Recall Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.

Event Detail

Event Id 64679
Product Type Devices
Status Ongoing
Recalling Firm Biomet U.K., Ltd.
City Bridgend, South Wales
State
Country GB
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-20
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
-
-