• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of April 24, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment.
Recall Number Z-1126-2013
Classification Class II
Code Info All units of these models.
Product Distributed Qty 176 units
Reason For Recall A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.

Event Detail

Event Id 64712
Product Type Devices
Status Ongoing
Recalling Firm Accuray Incorporated
City Sunnyvale
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-03-19
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.
-
-