• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of April 24, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description EndoGastric Solutions EsophyX2 Plus Device with Serosa Fastener and Accessories Part number C02033-01. Label reads in part "EsophyX2 2-LINK Device Qty: (1) Fastener Delivery System ***" Designed for the treatment of Gastroesophageal Reflux Disease (GERD), with erosive and non-erosive esophagitis
Recall Number Z-1136-2013
Classification Class II
Code Info Lots 401378, 401392, 401409, and 401416 Additional lots manufactured from 11/26/12 to 12/07/12 had a stop ship issued issued on 11/21/12 and the product was placed in quarantine (it was not distributed). This represents an additional 175 units from a total of 187 units manufactured during this time period. 12 units were used for bench testing. The lot numbers of units in quarantine are 401432, 401434, 401436, and 401441.
Product Distributed Qty 190
Reason For Recall EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians.

Event Detail

Event Id 64395
Product Type Devices
Status Ongoing
Recalling Firm Endogastric Solutions Inc
City Redmond
State WA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-02-04
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution including NC, VA, WI, TN, KY, IN, FL, MI, IA, IL, TX, LA, AR, HI, CA, NV, OR, and WA,
-
-