Product Detail

Product Description Software correction applies to customers who use the RT Therapist RTT4.2.108 and RT Therapist Connect RTT 4.2.108 on Siemens ARTISTE, ONCOR and PRIMUS systems. ARTISTE, ONCOR and PRIMUS family of linear accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Recall Number Z-1182-2013
Classification Class II
Code Info Syngo RT Therapist RTT 4.2.108; part number 8162815: Syngo ST Theratpist Connect RTT 4.2.108; part number 8168754
Product Distributed Qty 245 active devices
Reason For Recall Software update to fix multiple safety related issues.

Event Detail

Event Id 64944
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Martinez
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-04-04
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and internationally to Poland, India, Canada, Germany, United Kingdom, New Zealand, Belgium, Malaysia, Republic Korea, Ireland, Australia, Norway, Japan, Turkey, Croatia, Saudi Arabia, Colombia, P.R. China, Brazil, Thailand, Romania, Lebanon, Mexico, Italy, Sweden, Czech Republic, Russian Federation, and South Africa.
-
-