• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of March 19, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.
Recall Number Z-1201-2014
Classification Class II
Code Info serial numbers: 1289064, 1289069, 1289159, 1289259, 1289387, 1289650, 1289747, 1289826, 1289949, 1290088, 1290151, 1290333, 1290367, 1290480, 1290489, 1290845, 1290988, 1291042, 1291120, 1291233, 1291373, 1291378, 1291634, 1291777, 1291828, 1291933, 1292031, 1292114, 1292185, 1292185, 1292192, 1292391, 1292546, 1292597, 1292670, 1292674, 1292899, 1292960, 1293010, 1293039, 1293185, 1293190, 1293311, 1293379, 1293414, 1293687, 1293724, 1293771, 1293777, 1293953, 1293957, 1294093, 1294103, 1294171, 1294307, 1294362, 1294433, 1294516, 1294589, 1294727, 1294864, 1294999, 1295113, 1295137, 1295160, 1295186, 1295274, 1295391, 1295514, 1295622, 1295813, 1295871, 1296237, 1296241, 1296246, 1296484, 1296491, 1296582, 1296621, 1296650, 1296695, 1296806, 1297041, 1297046, 1297118, 1297231, 1297234, 1297372, 1297375, 1297394, 1297453, 1297484, 1297610, 1297655, 1297939, 1297943, 1298037, 1298210, 1298266, 1298270, 1298377, 1298407, 1298450, 1298457, 1298575, 1298619, 1298624, 1298704, 1298734, 1298797, 1298820, 1298879, 1298913, 1298952, 1298953, 1299037, 1299120, 1299144, 1299170, 1299241, 1299262, 1299285, 1299331, 1299393, 1299473, 1299512, 1299599, 1299655, 1299656, 1299732.
Product Distributed Qty 130
Reason For Recall There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Event Detail

Event Id 67516
Product Type Devices
Status Ongoing
Recalling Firm Mar Cor Purification
City Minneapolis
State MN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-27
Initial Firm Notification of Consignee or Public Telephone
Distribution Pattern nationwide. No distribution outside the US.
-
-