Product Detail

Product Description DICOM Export Format for Sum Dose with Velocity v3.0.0. A stand-alone software product that provides the oncology care specialists (physicians, physicists, dosimetrists, etc.) a means for comparison of medical imaging data from multiple DICOM conformant imaging modality sources. It allows the display, annotating, volume rendering, registration and fusing of medical images as an aid during use by diagnostic radiology, oncology, radiation therapy planning and other medical specialties.
Recall Number Z-1275-2014
Classification Class II
Code Info VelocityAIS 3.0.0
Product Distributed Qty 54 installs
Reason For Recall An anomaly was detected such that, under certain conditions, a summed dose created by Velocity replaces the calculated dose from an approved plan in Eclipse when importing the dose from Velocity.

Event Detail

Event Id 67615
Product Type Devices
Status Completed
Recalling Firm Velocity Medical Solutions, LLC
City Atlanta
State GA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-07-25
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - US (nationwide) including the states of AL, CA, CO, CT, DC, FL, GA, IL, IN, LA, MA, MD, MI, MS, NC, NY, PA, SC, TN, TX, WI, and the countries of Belarus, Italy and Switzerland.
-
-