Product Detail

Product Description i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.
Recall Number Z-1306-2014
Classification Class II
Code Info List number 03P78-25 :Lot number N13183
Product Distributed Qty 48, 450 US, 1400 OUS (49,850 catridges)
Reason For Recall Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.

Event Detail

Event Id 67513
Product Type Devices
Status Ongoing
Recalling Firm Abbott Point Of Care Inc.
City Princeton
State NJ
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-01-01
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and Internationally to Canada.
-
-