• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of June 5, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.
Recall Number Z-1419-2013
Classification Class II
Code Info Software version and build numbers 2.0.0.15, 2.5.1.89, and 3.0.0.251 (to determine if the version used is affected, bring up the About box in the RayStation application and check the build number reported0.
Product Distributed Qty Domestic: 70 units; Foreign: 27 units
Reason For Recall Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.

Event Detail

Event Id 65136
Product Type Devices
Status Ongoing
Recalling Firm RAYSEARCH LABORATORIES AB
City Stockholm
State
Country SE
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-02-15
Initial Firm Notification of Consignee or Public E-Mail
Distribution Pattern Worldwide Distribution - USA including Connecticut, Florida, Massachusetts, Missouri, New York, Ohio, and foreign distribution to(Canada, Italy, Australia, Germany, Netherlands, and Switzerland.
-
-