• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of April 16, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Recall Number Z-1440-2014
Classification Class I
Code Info Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005
Product Distributed Qty 3
Reason For Recall The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

Event Detail

Event Id 67677
Product Type Devices
Status Ongoing
Recalling Firm Draeger Medical Systems, Inc.
City Telford
State PA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-03-05
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern USA Nationwide Distribution in the states of WI and NY.
-
-