• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of June 12, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Comfort brand Ear-Loop Face Mask, Item #: 3100, Model Numbers: # 3100B, # 3100P, # 3100G, # 3100Y, # 3100W; Product is manufactured and distributed by Maytex Corporation, CA Ear-loop face mask for basic medical/dental procedures.
Recall Number Z-1451-2013
Classification Class II
Code Info All cases of item #: 3100 face mask with PO #. PO #, Date of receipt: 809-01, 2/5/08; 837-02, 7/29/08; 852-03, 10/10/08; 901-01, 2/3/09; 921-02, 6/8/09; 923-03, 7/8/09; 926-04, 8/26/09; 931-05, 11/10/09; 942-06, 12/16/09; 001-01, 2/17/10; 003-03, 4/12/10; 002-02 , 3/24/10; 003-03A, 7/23/10; 042-04, 11/29/10; 105-01, 1/31/11; 117-02, 4/29/11; 127/03, 7/8/11; 147-04, 12/16/11; 202-01, 3/8/2012; 224-02, 5/24/2012; 235-03, 9/12/2012; 253-04 , 12//192012
Product Distributed Qty 43,915 Cases
Reason For Recall Recent Lab testing conducted in March 2013 indicates that the product does not meet the standard of "Differential Pressure of only 1.6 Delta P (mm H2O/cm2)" as indicated on the box/package of the product; and also product manufactured between 2008 - 2012 cannot be guaranteed as Fluid Resistant.

Event Detail

Event Id 65160
Product Type Devices
Status Ongoing
Recalling Firm Maytex Corp
City Hayward
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-15
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution: USA (nationwide) including states of: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, MA, MD, ME, MI, MN, MO, MT, NE, NH, NJ, NM, NY, OK, OR, PA, PR, TN, TX, UT, WA, and WI; and country of: Mexico.
-
-