Product Detail

Product Description Part 510121 Disposable Infusion Pump, ...Elastomeric Pump 300 mL, 4mL/hr,...sterile.. found in the following kits GoPump Kit 510203-BP, Epidural catheter & BIOPATCH GoPump Kit 510204, 2.5" Fenestrated catheter GoPump Kit 510204 BP, 2.5 Fenestrated catheter & Biopatch GoPump Kit 510205BP, 5" Fenestrated catheter & Biopatch GoPump Kit 510551-BP, 2.5 Fenestrated catheter & Biopatch GoBlock kit 510608 Disposable Single Infusion Pump; Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.
Recall Number Z-1455-2013
Classification Class I
Code Info Pump Part 510121, Lot 11-100081 to 12-101406 found in: Go Pump Kit 510203-BP. Lot 12-101301 to 13-101538 GoPump Kit 510204, Lot 11-100505 to 11-1-665 GoPump Kit 510204 BP, Lot 11-100721 to 13-100680 GoPump Kit 510205BP, Lot 11-100722 to 13-100087 GoBlock kit 510608
Product Distributed Qty 33,543 in all kits
Reason For Recall Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the 300 mL (dual reservoir), 2 mL/hr GOPump (PN510076). To date, there have been 5 complaints received, 2 of which involved serious injury. There have been no patient deaths reported. The risk of patient toxicity due to the rapid influx of medication include seizure, dysrhythmia, and death.

Event Detail

Event Id 65192
Product Type Devices
Status Ongoing
Recalling Firm Symbios Medical Products, LLC
City Indianapolis
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-10
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern nationwide USA,
-
-