Product Detail

Product Description Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.
Recall Number Z-1457-2013
Classification Class I
Code Info Pump Part 510449, Lot 11-100271 - Lot 130100137 found in: GoBlock Kit 510472, Lot 11-100385 to 13-100354
Product Distributed Qty
Reason For Recall Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the 300 mL (dual reservoir), 2 mL/hr GOPump (PN510076). To date, there have been 5 complaints received, 2 of which involved serious injury. There have been no patient deaths reported. The risk of patient toxicity due to the rapid influx of medication include seizure, dysrhythmia, and death.

Event Detail

Event Id 65192
Product Type Devices
Status Ongoing
Recalling Firm Symbios Medical Products, LLC
City Indianapolis
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-10
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern nationwide USA,
-
-